Implementation of a Talbot-Lau interferometer in a clinical-like c-arm setup: A feasibility study

dc.contributor.authorHorn, Florian
dc.contributor.authorLeghissa, Martino
dc.contributor.authorKaeppler, Sebastian
dc.contributor.authorPelzer, Georg
dc.contributor.authorRieger, Jens
dc.contributor.authorSeifert, Maria
dc.contributor.authorWandner, Johannes
dc.contributor.authorWeber, Thomas
dc.contributor.authorMichel, Thilo
dc.contributor.authorRiess, Christian
dc.contributor.authorAnton, Gisela
dc.date.accessioned2019-05-10
dc.date.available2019-04-18
dc.date.created2018
dc.date.issued2019-05-10
dc.description.abstractX-ray grating-based phase-contrast imaging has raised interest regarding a variety of potential clinical applications, whereas the method is feasible using a medical x-ray tube. Yet, the transition towards a clinical setup remains challenging due to the requirement of mechanical robustness of the interferometer and high demands applying to medical equipment in clinical use. We demonstrate the successful implementation of a Talbot-Lau interferometer in an interventional c-arm setup. The consequence of vibrations induced by the rotating anode of the tube is discussed and the prototype is shown to provide a visibility of 21.4% at a tube voltage of 60 kV despite the vibrations. Regarding clinical application, the prototype is mainly set back due to the limited size of the field of view covering an area of 17 mm × 46 mm. A c-arm offers the possibility to change the optical axis according to the requirements of the medical examination. We provide a method to correct for artifacts that result from the angulation of the c-arm. Finally, the images of a series of measurements with the c-arm in different angulated positions are shown. Thereby, it is sufficient to perform a single reference measurement in parking position that is valid for the complete series despite angulation.en
dc.identifier.citationScientific Reports 8 (2018). <https://doi.org/10.1038/s41598-018-19482-z>
dc.identifier.doihttps://doi.org/10.1038/s41598-018-19482-z
dc.identifier.opus-id10867
dc.identifier.urihttps://open.fau.de/handle/openfau/10867
dc.identifier.urnurn:nbn:de:bvb:29-opus4-108678
dc.language.isoen
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/deed.de
dc.subjectRadiography
dc.subjectX-rays
dc.subject.ddcDDC Classification::5 Naturwissenschaften und Mathematik :: 50 Naturwissenschaften :: 500 Naturwissenschaften und Mathematik
dc.titleImplementation of a Talbot-Lau interferometer in a clinical-like c-arm setup: A feasibility studyen
dc.typearticle
dcterms.publisherFriedrich-Alexander-Universität Erlangen-Nürnberg (FAU)
local.journal.titleScientific Reports
local.journal.volume8
local.sendToDnbfree*
local.subject.fakultaetNaturwissenschaftliche Fakultät
local.subject.sammlungUniversität Erlangen-Nürnberg / Von der FAU geförderte Open Access Artikel / Von der FAU geförderte Open Access Artikel 2018